◆ Power Supply
220V\110V±10%,50/60Hz
◆ Operation Pressure
100-10000 mbar (1.5-145psi)
◆ Min inlet pressure
100 mbar (1.5psi)
◆ Dimension(mm)
400(Depth)*240(Width)*260(Height1)*220(Height2)
◆ Function
Manual Bubble point Test;Basic Bubble Point Test;Extensive Bubble Point Test; Pressure Holding Test;Diffusion FLow Test;Water Intrusion Test;Ultrafitration Membrane Test
◆ Test Accuracy
Upstream Volume Test:≤± 4%;Bubble Point Test:≤± 50mbar ;Diffusion Flow Test:≤± 4% ;Water Immersion:≤± 0.01ml
◆ Test Scope
BP:100-8000mbar DF:1-600ml/min WI:0.01-100ml/min:0.01-100ml/min
◆ Applied Range
Symmetric and asymmetric membrane test, needle filter, filter bag, flat filter, filter cartridge (big filter volume ), ultrafiltration membrane, ultrafiltration column, a variety of irregular filters
◆ Authority Management
User Name and Password to Login ,electronic signature
◆ Reservation Solution
120 sets
◆ Print Function
Built-in printer;
◆ Records Copy
Support USB disk to export (Include test curve)
◆ Display Screen
7 inch, color touch screen
◆ Serial Connection
RS232/USB
◆ Language
Chinese/English
◆ Work Mode
Online/Offline
◆ Other functions
One/two-dimension cods ,Scanning gun;Lithium battery(customization)
◆ Applicable Environment
Above D level
◆ Weight
8KG
◆ Signal output
(4-20)MA,RS485,12V Alarm output (customization)
◆ 重量
7.5kg
◆ Dimension
400*260*240mm